Evaluation of the MAGLUMI HIV Ab/Ag combi test for the detection of HIV infection
Virology Journal volume 21, Article number: 290 (2024)
Abstract
Background
Human immunodeficiency virus (HIV) infection screening and diagnosis are critical to control the HIV epidemic. Testing for anti-HIV antibodies (Ab) and antigens (Ag) in blood samples is the first step to screen people who have been potentially exposed to the virus. This study aimed to evaluate the performance of the MAGLUMI HIV Ab/Ag Combi for detection of HIV antibodies and antigens.
Methods
We used residual samples to assess the diagnostic specificity and sensitivity of the MAGLUMI HIV Ab/Ag Combi retrospectively. All samples that met the test criteria were tested with the MAGLUMI HIV Ab/Ag Combi according to manufacturer’s instruction. Results of the MAGLUMI HIV Ab/Ag Combi were compared with the Architect HIV Ag/Ab Combo test.
Results
The specificity of the MAGLUMI HIV Ab/Ag Combi was 99.85% in 5,057 unselected blood donors and 100.00% in 213 hospitalized patient samples, respectively. The sensitivity of the Test in 614 HIV-1 Ab, HIV-1 Ag or HIV-2 Ab positive samples was 100.00%. Seroconversion sensitivity from results of 30 panels was comparable between the MAGLUMI HIV Ab/Ag Combi and the Architect assay.
Conclusions
The reactivity of the MAGLUMI HIV Ab/Ag Combi test is comparable to the Architect HIV Ag/Ab Combo assay.
Introduction
In 2023, about 1.3 million new HIV infections occurring and approximately 630 thousand people died from HIV-related causes [1, 2]. Even though there is no cure or effective vaccine valid for HIV infection [3], antiretroviral therapy (ART) is able to turn HIV infection into a chronic disease, provide longer lives for patients and reduce HIV transmission [4]. The challenge is that HIV-infected individuals are unaware of their status until the later stages and fail to receive timely treatment [5]. This is due to the latency of HIV infection that in the first few weeks after being infected people may have no symptoms or only have an influenza-like illness, following with a chronic phase persisting no symptoms or mild ones for years [5]. Without treatment, people with HIV infection can develop severe illnesses including tuberculosis, cryptococcal meningitis, pneumocystis pneumonia and cancers [6,7,8,9]. Therefore, the detection and management of early HIV infection is critical to improve patient health, reduce their additional medical burden and control the risk of onward viral transmission [10]. However, the diagnosis rate of HIV infections is still far from satisfactory, about 5.5 million people did not know they were living with HIV in 2022 [2].
The US Centers for Disease Control and Prevention (CDC) recommends that everyone aged between 13 and 64 get tested for HIV at least once as part of routine health care. People with certain risk factors should get tested at least once a year [11]. Sexually active Men who have sex with men (MSM) were recommended to get more frequent testing (every 3 to 6 months) [11]. Testing for both of antibodies (Ab) and antigens (Ag) in serum and plasma specimens is considered to be the first step to screen the people who have been potentially exposed to HIV [12, 13]. Additionally, a confirmatory HIV test and/or subsequent HIV RNA testing are required for specimens with a positive result or a suspected very early infection [12, 13]. Although the HIV RNA test is useful for a recent high-risk HIV exposure, confirming a result from another HIV test and monitoring viral load [14, 15], HIV RNA test is not recommended to initially screen HIV infection due to its expensive fees with only slightly shortened window period (Supplementary Figure S1) [16, 17].
The current commercial tests for HIV infection diagnosis widely used mainly belong to the Ab or Ag/Ab tests. Ag/Ab assays detect HIV antibodies together with the p24 antigen, allowing for more sensitive detection of recent HIV infection and reduction of the window period by comparing with Ab assays, which only detect antibodies [18]. In recent years, several new automated HIV Ag/Ab assays have been brought to market, including Roche Elecsys HIV Duo assay, Beckman Access HIV Ag/Ab combo, Beckman Access HIV combo V2 assay and Snibe MAGLUMI HIV Ab/Ag Combi (MAGLUMI) [19,20,21]. MAGLUMI is based on the double-antigen sandwich principle and uses recombinant HIV-1/HIV-2 antigens to detect HIV-1 and/or HIV-2 antibodies, in the meanwhile, anti-HIV-1 p24 monoclonal antibodies are used to detect HIV-1 antigens. In this study, we evaluated the clinical performance of the MAGLUMI to verify whether the test is suited for HIV infections.
Materials and methods
MAGLUMI HIV Ab/Ag combi test
MAGLUMI is a two-step sandwich chemiluminescence immunoassay. The first step: The sample, the magnetic microbeads and N-(4-Aminobutyl)-N-ethylisoluminol (ABEI) are mixed thoroughly and incubated. Here the magnetic microbeads coated with anti-HIV-1 p24 monoclonal antibodies and HIV-1/HIV-2 recombinant antigens (recombinant gp41 of HIV-1 and gp36 of HIV-2). The ABEI labeled with anti-HIV-1 p24 monoclonal antibodies. The antibodies to HIV-1 and/or HIV-2 present in the sample bind to the HIV-1 and HIV-2 recombinant antigens to form a complex, and the HIV-1 p24 antigens present in the sample bind to the anti-HIV-1 p24 monoclonal antibodies. The second step: After washing, ABEI labeled with HIV-1 and HIV-2 recombinant antigens (recombinant gp41 of HIV-1 and gp36 of HIV-2) are added, and bind to the complex. Following another wash cycle, the rest unbound materials are removed, the Starter 1 + 2 are added to initiate a flash chemiluminescent reaction. The light signal is measured by a photomultiplier as relative light units (RLUs), which is indicative of the concentration of HIV p24 antigens and antibodies against HIV-1 and/or HIV-2 present in the sample. Biotin is excluded in the assay design and there is no interference when biotin concentration is not exceeding 50 µg/mL. All samples in this study were tested with MAGLUMI according to manufacturer’s instruction.
Blood donor and patient samples
A total of 5,057 blood donor samples (serum or plasma) with negative status for the Architect HIV Ag/Ab Combo (Architect) were collected from two blood donation centers in Germany (Table 1). These samples were directly used for assessing the performance of diagnostic reagents and the donation centers did not provide the HIV RNA results for this study, although we know that HIV-1 RNA test was mandatory in regular blood donation centers in Germany. All samples were tested with MAGLUMI and the samples with a result greater than or equal to 1.00 AU/mL (≥ 1.00 AU/mL) were considered to be positive and retests would be performed.
Blood donor and HIV positive samples used in this study were all residual samples with known test values. A total of 614 HIV-positive samples were included in the sensitivity calculation, 459 HIV-1 Ab, 105 HIV-2 Ab and 50 HIV-1 Ag samples (Table 1). Samples were either initially tested with Architect, Abbott PRISM or BioMérieux VIDAS HIV-DUO. Initially reactive samples were confirmed for antibody presence by Bio-Rad Geenius HIV-1/2 Confirmatory Assay. The presence of HIV-1 antigen p24 for these samples were confirmed by DiaSorin LIAISON XL Murex HIV Ab/Ag, BIO-RAD BioPlex 2200 HIV Ag/Ab assay or Roche Elecsys HIV-1 Ag. Among the HIV-positive samples, 436 HIV-1 Ab, 55 HIV-2 Ab and all 50 HIV-1 Ag samples have test results of Architect. Furthermore, the HIV-positive status of all HIV-1 Ab and Ag positive samples were confirmed by HIV-1 RNA test on Abbott m2000 RealTime System. Among these HIV positive samples, 270 HIV-1 and 12 HIV-2 Ab positive samples were with known subtype respectively. Specifically, 10 HIV-1 subtypes and 13 HIV-1 CRFs were included in these HIV-1 positive samples with known subtype.
Samples from 213 consecutive hospitalized patients were used to evaluate the analytical specificity of the test (Table 1). All these patients have a clinical diagnosis of other diseases and no history of clinical symptoms and epidemiology related to HIV infection. Samples of these patients were selected and then tested by both MAGLUMI and Architect.
Seroconversion sensitivity
30 commercially available HIV seroconversion panels representing a total of 356 samples were tested (Table 1). The panels were: SCP-HIV1-002, SCP-HCV1-007 (Biomex GmbH, Heidelberg, Germany), HIV 6244, HIV 6248, HIV9011, HIV9012, HIV9013, HIV9016, HIV9018, HIV9020, HIV9021, HIV9022, HIV9023, HIV9030, HIV9031, HIV9034, HIV9076, HIV9077, HIV9079, HIV9089, HIV9096, HIV12008 (ZeptoMetrix, Buffalo, NY, USA), PRB945, PRB955, PRB963, PRB966, PRB968, PRB969, PRB973, 0600–0271 (SeraCare Inc., Milford, MA, USA). The panels were evaluated between MAGLUMI and Architect to assess how early two assays could detect infection. Numbers of positive samples and the first positive day of each panel were determined and compared to evaluate the sensitivity on seroconversion. The shorter undetectable days (number of days between the first Ab/Ag positive day and the first blood day) or the more detectable panel samples mean the test detected HIV Ab/Ag earlier. Data of Architect were collected from the seroconversion panel datasheets.
Endogenous sample interference
In total 110 samples with potentially cross reacting substance/ agents including samples generated from patients with auto-immune diseases, pregnant women, patients with hyper IgG/IgM, flu patients, dialysis patients and patients with infectious diseases were evaluated (Table 1). 27 samples with interfering substances including hemolytic, lipemic and icteric were assessed (Table 1).
Analytical sensitivity in international standard and national reference
The twofold dilution series of the NIBSC HIV-1 p24 antigen WHO international standard (code: 90/636) were tested in parallel on MAGLUMI and Architect (Table 1). HIV-1 Ab (Lot: 370,045–201901) and p24 antigen (Lot: 220,015–201906) China national reference panels were also tested on MAGLUMI (Table 1).
Statistical analysis
Calculation of proportions and corresponding Wilson 95% confidence intervals were performed using Analyse-IT v4.81 (Leeds, United Kingdom). Analysis of differences between various groups of MAGLUMI AU values and correlation analysis of the results obtained on HIV-1 Ab, HIV-2 Ab and HIV-1 Ag positive samples were performed by GraphPad Prism Software (San Diego, CA, USA). Differences between various groups of MAGLUMI AU values were determined using the Kruskal–Wallis test. Linear correlations of the results obtained on HIV-1 Ab, HIV-2 Ab and HIV-1 Ag positive samples between MAGLUMI and Architect were determined by the Pearson correlation coefficient.
Results
Diagnostic specificity and sensitivity
Among 5,270 Ag/Ab negative samples from 5,057 blood donors and 213 hospitalized patients, 8 gave presumed false positive results on MAGLUMI. This resulted in a diagnostic specificity of 99.85% (5,262/5,270) (Table 2). After centrifuged and transferred the clarified specimen to a second tube for retesting, seven samples were repeatedly below the MAGLUMI cutoff (Supplementary Table S1).
All 459 HIV-1 Ab and 105 HIV-2 Ab positive samples included 282 samples with a known HIV genotype as well as 50 HIV-1 Ag positive samples were found positive with MAGLUMI which presented a 100.00% diagnostic sensitivity (Table 3). Both MAGLUMI and Architect are chemiluminescent immunoassays and the relative light units correlate with the concentration of the measured substance in a certain range, which gives them the potential for determining recent infection based on different values [22]. So we did correlation analysis of the results with specific test values obtained by both assays on HIV-1 Ab, HIV-2 Ab and HIV-1 Ag positive samples respectively. We found a very poor correlation of HIV antibody values between MAGLUMI and Architect (Suppl. Figure S2). However, a good correlation of HIV-1 Ag values was observed (Suppl. Figure S2).
Seroconversion sensitivity
In this study, 30 commercially available HIV seroconversion panels were tested to evaluate seroconversion sensitivity of MAGLUMI. For each panel, numbers of positive samples and delayed days of the first positive days based on serological test results were compared (Table 4, supplementary Table S2). Comparing with Architect, MAGLUMI detected one more positive sample in two panels and the same samples in the rest panels. Therefore, MAGLUMI detected slightly more panel samples (125 of 356) compared with Architect (123 of 356) on 30 seroconversion panels (Table 4). Total delayed days since the first blood draw of 30 panels tested were slightly fewer in MAGLUMI (1,105 days) compared with Architect (1,115 days), which indicated that MAGLUMI detected HIV infection as early as Architect (Supplementary Table S2). It is worth noting that one panel (SCP-HIV1-007) starts with a negative sample, seroconverts and turns back to negative, and the same effect was seen in both assays. 114 samples from 27 seroconversion panels were defined as early seroconversion HIV samples. Early seroconversion samples were defined in the Common Technical Specifications (2002/364/EC and amendments) as p24 antigen and/or HIV RNA positive, not recognized by all of the antibody screening tests, and indeterminate or negative confirmatory assays [23]. Among these early seroconversion HIV samples, 60 and 58 were detected as positive with MAGLUMI and Architect, respectively (Supplementary Table S3). In conclusion, based on numbers of positive samples and delayed days of the first positive days, MAGLUMI performed comparable on seroconversion sensitivity compared with Architect.
Analytical specificity
All 137 samples with potentially cross reacting substances/agents or interfering substances were negative with MAGLUMI (Table 5). The analytical specificity on 137 samples with potential cross reacting/interfering substances was 100.00% (137/137).
Analytical sensitivity
Both MAGLUMI and Architect detected the HIV-1 p24 antigen international standard with a similar sensitivity (Table 6). In this evaluation, the analytical sensitivity of both assays was 0.78 IU/mL which is in compliance with the common technical specification (CTS): ≤ 2 IU/mL. In addition, MAGLUMI detected both the HIV-1 Ab and p24 antigen China national reference panels with an acceptable analytical sensitivity (Supplementary Table S4). In p24 antigen national reference evaluation, the analytical sensitivity of the assay was 1.153 IU/mL which was in compliance with the CTS: ≤ 2.5 IU/mL. The obtained results of HIV-1 Ab (Limit of Detection) LoD and HIV-1 p24 antigen linearity as well as LoD were summarized in Supplementary Table S5.
Discussion
This article reports the sensitivity and specificity of MAGLUMI for use on the MAGLUMI series fully-auto chemiluminescence immunoassay analyzer. The results showed that MAGLUMI had a specificity of 99.85% (5,262/5,270) in samples from blood donors and hospitalized patients, and a sensitivity of 100.00% (614/614) in HIV positive samples. The test also possessed a comparable seroconvertion sensitivity with Architect. In addition, we observed a high analytical specificity and sensitivity by evaluating the influence of factors including interfering substances, international standard, national reference, complement interference and serum to plasma equivalence. Based on the Common Technical Specifications 2002/364/EC and amendments, it can be concluded that MAGLUMI has a reasonably good specificity and sensitivity.
Among samples used to assess diagnostic specificity, eight gave presumed false positive results and one of them remained positive result after centrifugation. Actually, Fibrin were observed in all eight presumed false positive samples, emphasizing on the correlation between sample condition, necessary reparation for analysis and reliable results. One limitation of this study is that lack of information on HIV RNA status of blood donors, hospitalized patients and HIV-2 Ab samples. Therefore, the presumed false positive results detected among these samples have not been confirmed.
HIV is a highly variable virus characterised by extensive genetic heterogeneity [24, 25], which brings a challenge to the development of vaccines and pan-genotypic treatments [26], and also requires diagnostic tests applicable to the detection of all genotypes. HIV can be divided into 2 lineages, HIV-1 and HIV-2 [27]. HIV-1 is the predominant lineage in the global epidemic accounts for 95% of infections. Based on the differences in nucleic acid sequence of env gene, HIV-1 is divided into 4 groups (M, N, O, and P), among which the main group is M group including at least 9 subtypes (A-D, F–H, J, K) and various circulating recombinant forms (CRFs) [27,28,29]. HIV-1 M group infection is the most prevalent globally, subtypes like C, B, A and CRFs like CRF02_AG, CRF01_AE are widely distributed worldwide and account for a great proportion of the totality of HIV cases [24]. In this study, 270 HIV-1 Ab positive samples with 10 subtypes and 13 CRFs comprehensively covered most of the major subtypes. In addition, 12 HIV-2 Ab positive samples with subtype A or B were evaluated in this study. However, the detection rates of other HIV gene subtypes using this assay need to be verified.
MAGLUMI is an Ag/Ab chemiluminescence immunoassay (CLIA). The CLIA has been greatly developed in the past several decades and extensively used in diagnosis of infectious disease including HIV, hepatitis B and hepatitis C virus [30,31,32]. In some high-volume clinical laboratories, the CLIA is replacing the traditional EIA method in some high-volume clinical laboratories due to its technical simplicity, full automation and greater positive predictive value [33]. Recombinant protein gp41 of HIV-1, gp36 of HIV-2 and anti-HIV-1 p24 monoclonal antibody are widely used for the development with HIV Ag/Ab assays [19,20,21]. Besides, Beckman Access HIV combo V2 assay allows extra detection of HIV-2 p26 Ag [21]. It is noteworthy that a recent study showed the new Access HIV Ag/Ab combo assay exhibited better results of viral lysates and early detection on seroconversion panels compared with Architect and Roche Elecsys HIV Duo [20]. Therefore, to evaluate performance of MAGLUMI more comprehensively, it is necessary to compare its performance with more HIV Ag/Ab assays, especially the more sensitive assays approved in recent years.
Several studies claimed that very low S/CO values predict false-positive results and the false positive rate is reduced when S/CO increases [34,35,36]. In addition, current status of HIV infection could also be predicted by the signal-to-cutoff (S/CO) ratio derived from HIV Ag/Ab tests [22]. However, variable optimal S/CO values generated from different HIV Ag/Ab tests should be taken into account when predicting true HIV infection status. We observed a good correlation of positive Ag results, but poor correlations of both HIV-1 and HIV-2 Ab between MAGLUMI and Architect even though both tests performed high specificity and sensitivity. One possible reason could be differential material ratios the two tests used.
The Global Health Sector Strategies (GHSS) promote the disease-specific goal to end AIDS by 2030 [37]. To achieve this ambitious goal, by 2025, 95% of all people living with HIV should know their HIV status, 95% of all people with diagnosed HIV infection should be receiving sustained antiretroviral therapy and 95% of all people on treatment should have viral suppression by 2025 [2, 38]. However, only 86% of all people living with HIV knew their HIV status in 2022 [2]. “The first 95” target is fundamental to these ambitious goals, this way the infection can be timely monitored, controlled and treated to enable people living with HIV to avoid the further development of serious complications. MAGLUMI showed a comparable reactivity with Architect, which makes it have the potential to contribute to reaching “The first 95” target.
Availability of data and materials
No datasets were generated or analysed during the current study.
References
World Health Organization. HIV and AIDS. Available online: https://www.who.int/news-room/fact-sheets/detail/hiv-aids (accessed on 22 July 2024).
UNAIDS. Global HIV and AIDS statistics — fact sheet. Available online: https://www.unaids.org/en/resources/fact-sheet.
Landovitz RJ, Scott H, Deeks SG. Prevention, treatment and cure of HIV infection. Nat Rev Microbiol. 2023;21(10):657–70.
Menéndez-Arias L, Delgado R. Update and latest advances in antiretroviral therapy. Trends Pharmacol Sci. 2022;43(1):16–29.
Daar ES, Pilcher CD, Hecht FM. Clinical presentation and diagnosis of primary HIV-1 infection. Curr Opin HIV AIDS. 2008;3(1):10–5.
Bell LCK, Noursadeghi M. Pathogenesis of HIV-1 and mycobacterium tuberculosis co-infection. Nat Rev Microbiol. 2018;16(2):80–90.
Williamson PR, Jarvis JN, Panackal AA, Fisher MC, Molloy SF, Loyse A, et al. Cryptococcal meningitis: epidemiology, immunology, diagnosis and therapy. Nat Rev Neurol. 2017;13(1):13–24.
Fei MW, Sant CA, Kim EJ, Swartzman A, Davis JL, Jarlsberg LG, et al. Severity and outcomes of pneumocystis pneumonia in patients newly diagnosed with HIV infection: an observational cohort study. Scand J Infect Dis. 2009;41(9):672–8.
Yarchoan R, Uldrick TS. HIV-associated cancers and related diseases. N Engl J Med. 2018;378(11):1029–41.
Smith MK, Rutstein SE, Powers KA, Fidler S, Miller WC, Eron JJ Jr, et al. The detection and management of early HIV infection: a clinical and public health emergency. J Acquir Immune Defic Syndr. 2013;63(2):S187–99.
Centers for Disease Control and Prevention. Getting Tested for HIV. Available online: https://www.cdc.gov/hiv/testing/index.html. (accessed on 18 January 2024).
Branson BM. HIV diagnostics: current recommendations and opportunities for improvement. Infectious dis clin of N Am. 2019;33(3):611–28.
World Health Organization. (2019). Consolidated guidelines on HIV testing services for a changing epidemic: policy brief. Available online: Consolidated guidelines on HIV testing services for a changing epidemic (who.int) (accessed on 27 November 2019).
De Souza MS, Phanuphak N, Pinyakorn S, Trichavaroj R, Pattanachaiwit S, Chomchey N, et al. Impact of nucleic acid testing relative to antigen/antibody combination immunoassay on the detection of acute HIV infection. AIDS. 2015;29(7):793–800.
Newman H, Hardie D. HIV-1 viral load testing in resource-limited settings: challenges and solutions for specimen integrity. Rev Med Virol. 2021;31(2):e2165.
Saag MS. HIV infection - screening, diagnosis, and treatment. N Engl J Med. 2021;384(22):2131–43. https://doiorg.publicaciones.saludcastillayleon.es/10.1056/NEJMcp1915826.
Nasrullah M, Wesolowski LG, Meyer WA 3rd, Owen SM, Masciotra S, Vorwald C, et al. Performance of a fourth-generation HIV screening assay and an alternative HIV diagnostic testing algorithm. AIDS. 2013;27(5):731–7.
Vallefuoco L, Mazzarella C, Portella G. Fourth generation assays for HIV testing. Expert Rev Mol Diagn. 2016;16(7):723–32.
Krasowski MD, Wier D, Smith S, Riedel A, Lauseker-Hao Y, Kelner M, et al. Real-world clinical performance evaluation of a fourth-generation HIV antigen/antibody differentiation test. J Appl Lab Med. 2021;6(6):1417–32. https://doiorg.publicaciones.saludcastillayleon.es/10.1093/jalm/jfab069.
Lemée V, Gréaume S, Gautier J, Dzamitika SA, Coignard C, Jortani SA, et al. Performance evaluation of the new Access HIV Ag/Ab combo assay on the DxI 9000 access immunoassay analyzer. J Clin Virol. 2024;174:105712. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.jcv.2024.105712.
Guiraud V, Ciczora Y, Cardona M, Defer C, Gréaume S, Nogues D, et al. Sensitivity and specificity of the new Bio-Rad HIV screening test, access HIV combo V2. J Clin Microbiol. 2024;62(5):e0009524. https://doiorg.publicaciones.saludcastillayleon.es/10.1128/jcm.00095-24.
Curtis KA, Rudolph DL, Pan Y, Delaney K, Anastos K, DeHovitz J, et al. Evaluation of the abbott architect HIV Ag/Ab combo assay for determining recent HIV-1 infection. PLoS ONE. 2021;16(7):e0242641.
Commission Decision of 27 November 2009 amending Decision 2002/364/EC on common technical specifications for in vitro diagnostic medical devices (2009/886/EC), http://eur-lex.europa.eu/LexUriServ/LexUriServ.do?uri=OJ:L:2009:318:0025:0040:EN:PDF
Hemelaar J, Elangovan R, Yun J, Dickson-Tetteh L, Fleminger I, Kirtley S, et al. Global and regional molecular epidemiology of HIV-1, 1990–2015: a systematic review, global survey, and trend analysis. Lancet Infect Dis. 2019;19(2):143–55.
Paraskevis D, Hatzakis A. Global molecular epidemiology of HIV-1: the chameleon challenge. Lancet Infect Dis. 2019;19(2):114–5.
Chen Z, Julg B. Therapeutic vaccines for the treatment of HIV. Transl Res. 2020;223:61–75.
Williams A, Menon S, Crowe M, Agarwal N, Biccler J, Bbosa N, et al. Geographic and population distributions of human immunodeficiency virus (HIV)–1 and HIV-2 circulating subtypes: a systematic literature review and meta-analysis (2010–2021). J Infect Dis. 2023;228(11):1583–91.
Bbosa N, Kaleebu P, Ssemwanga D. HIV subtype diversity worldwide. Curr Opin HIV AIDS. 2019;14(3):153–60.
Los Alamos National Laboratory. HIV circulating recombinant forms (CRFs). Available online: https://www.hiv.lanl.gov/components/sequence/HIV/crfdb/crfs.comp. (accessed on 21 July 2023).
Xiao Q, Xu C. Research progress on chemiluminescence immunoassay combined with novel technologies. TrAC - Trends in Anal Chem. 2020;124:115780.
Fei CR, Ye AQ, Zhang J. Evaluation of different methods in determination of low level HBsAg. Zhejiang Da Xue Xue Bao Yi Xue Ban. 2011;40(4):436–9.
Li T, Zhang H, Fang Z, Yin J, Rao W. Clinical performance of the MAGLUMI Anti-HCV (CLIA) test for detection of hepatitis C virus antibodies. J Virol Methods. 2023;319:114770.
Kim S, Kim JH, Yoon S, Park YH, Kim HS. Clinical performance evaluation of four automated chemiluminescence immunoassays for hepatitis C virus antibody detection. J Clin Microbiol. 2008;46(12):3919–23.
White DAE, Anderson ES, Basham K, Ng VL, Russell C, Lyons MS, et al. Clinical utility of the signal-to-cutoff ratio of reactive HIV antigen/antibody screening tests in guiding emergency physician management. J Acquir Immune Defic Syndr. 2022;89(3):332–9.
Marson KG, Marlin R, Pham P, Cohen SE, Jones D, Roemer M, et al. Real-world performance of the new US HIV testing algorithm in medical settings. J Clin Virol. 2017;91:73–8.
Chacón L, Mateos ML, Holguín Á. Relevance of cutoff on a 4th generation ELISA performance in the false positive rate during HIV diagnostic in a low HIV prevalence setting. J Clin Virol. 2017;92:11–3.
World Health Organization. Global health sector strategies on, respectively, HIV, viral hepatitis and sexually transmitted infections for the period 2022–2030. Available online: https://www.who.int/publications-detail-redirect/9789240053779 (accessed on 18 July 2022).
Frescura L, Godfrey-Faussett P, Feizzadeh AA, El-SadrW SO, Ghys PD. On and behalf of the 2025 testing treatment target working group achieving the 95 95 95 targets for all: a pathway to ending AIDS. PLoS ONE. 2022;17(8):e0272405.
Acknowledgements
We thank Dr. Heike Lukhaup for her coordination of the trail, Stefanie Von Cleaf for her assistance in preparing samples and Alisa Schäfer for performing the experiments. They all come from Biomex GmbH Heidelberg.
Funding
None declared.
Author information
Authors and Affiliations
Contributions
Conceptualization, C.Z. and T.L.; methodology, C.W., J.R., Z.F. and J.Y.; formal analysis, C.W., J.R., Z.F., H.Z. and J.Y.; writing—original draft preparation, H.Z.; writing—review and editing, C.W., J.R., Z.F., C.Z. and T.L.; supervision, Z.F.; project administration, Z.F.; funding acquisition, C.Z. and T.L.
Corresponding authors
Ethics declarations
Ethics approval and consent to participate
The clinical performance study was conducted by a third-party organization (laboratories of Biomex GmbH, Heidelberg, Germany) and carried out in accordance with the ethical principles laid down in the Declaration of Helsinki. The samples used in this study were all residual samples with a general authorization for ethical approval and subjects signed broad informed consent.
Consent for publication
All authors give consent for the publication of the data in this study.
Competing interests
The authors declare no competing interests.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Wang, C., Rao, J., Fang, Z. et al. Evaluation of the MAGLUMI HIV Ab/Ag combi test for the detection of HIV infection. Virol J 21, 290 (2024). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12985-024-02565-x
Received:
Accepted:
Published:
DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12985-024-02565-x