Table 4 Seroconversion sensitivity: comparison of MAGLUMI HIV Ab/Ag Combi and Architect HIV Ag/Ab Combo with 30 seroconversion panels

From: Evaluation of the MAGLUMI HIV Ab/Ag combi test for the detection of HIV infection

Panel ID

Panel samples

MAGLUMI HIV Ab/Ag Combi

Architect HIV Ag/Ab Combo

PRB945

6

3

3

PRB955

5

4

4

PRB963

7

2

2

PRB966

10

3

3

PRB968

10

4

4

PRB969

10

4

4

PRB973

4

2

2

HIV6244*

13

1

1

HIV6248

7

2

2

HIV9011

11

2

2

HIV9012

8

3

3

HIV9013

7

2

1

HIV9016

10

2

2

HIV9018

11

3

3

HIV9020

22

3

3

HIV9021

17

4

4

HIV9022*

8

1

1

HIV9023

22

3

3

HIV9030

16

3

3

HIV9031

19

4

3

HIV9034

13

3

3

HIV9076*

9

3

3

HIV9077*

24

13

13

HIV9079

25

17

17

HIV9089

6

3

3

HIV9096

6

5

5

HIV12008

13

5

5

SCP-HIV1-002

20

11

11

SCP-HIV1-007**

9

4

4

0600–0271

8

6

6

Total on 30 panels

356

125

123

  1. * HIV6244-14, -15, HIV9022-9, HIV9076-07, HIV-9077–23, -26, -27, -28 and -29 were exhausted and therefore not tested. For both tests the same panel samples were compared
  2. ** SCP-HIV1-007 starts with a negative sample, seroconverts and turns back to negative
  3. HIV, human immunodeficiency virus; Ab, antibodies; Ag, antigens